System Warnings
DO NOT update your Tandem insulin pump before reading the User Guide. Incorrect use of the Tandem Device Updater or failure to follow the appropriate instructions can lead to an inoperable pump. If you have questions or need further clarification on the
Tandem Device Updater or pump use, call Customer Technical Support at (877) 801-6901.
DO NOT update your pump while your infusion set is connected to your body. Always ensure that the infusion set is disconnected from your body before updating your pump software.
DO suspend all pump insulin delivery prior to using the Tandem Device Updater.
DO check your blood glucose (BG) prior to suspending delivery and be sure to treat high or low BG levels as directed by your healthcare professional prior to update.
DO be prepared to inject insulin with an alternative method in case you encounter any issues while updating your pump. Failure to have an alternative method of insulin delivery can lead to very high blood glucose or Diabetic Ketoacidosis (DKA).
DO complete the training appropriate to your device and feature set before starting to use the updated software. Failure to complete necessary training could result in serious injury or death.
DO NOT rely on the Insulin On Board (IOB) displayed on your pump for a period of time after an update. Your IOB will be reset to zero during the update process. Since the calculated bolus amount relies on IOB, it could prompt you to deliver more insulin
than needed and result in hypoglycemia. Consult with your healthcare professional for how long you need to wait after an update before you can rely on the IOB calculation.
DO confirm your pump’s personal settings, date, and time are correct after the update. Incorrect settings can result in over delivery or under delivery of insulin. Consult with your healthcare professional as needed to establish appropriate settings.